VIKING-EC500 Benchtop Conductivity Meter — COLO.Science
VIKING-EC500 benchtop conductivity meter is built for laboratories that demand high accuracy, intuitive operation, and robust data handling.
With a 7-inch touchscreen, up to 5-point calibration, temperature compensation, and GLP-compliant logging, it delivers a complete platform for conductivity measurement.
Conductivity range: 0.000 µS/cm to 2000 mS/cm · Accuracy: ±0.5% FS · Data storage: 1000 sets per parameter
Key Features
- 7-inch high-contrast color LCD touchscreen
- Intelligent OS with user, calibration, method and data management (GLP)
- Calibration up to 1–5 points with automatic standard recognition
- Automatic or manual temperature compensation (ATC/MTC)
- Multiple reading modes: Auto-Read, Timed-Read, Continuous-Read
- USB and RS-232 communication; on-meter data analysis (review/compare/recalculate)
- IP54 splash protection; reset to factory defaults
Ideal for water quality labs, research centers, and QC environments where traceable, repeatable results are critical.
Technical Specifications
Model | VIKING-EC500 |
---|---|
Measured parameters | EC / Resistivity / TDS / Salinity / Temperature |
Conductivity range | 0.000 µS/cm – 2000 mS/cm |
Conductivity resolution | 0.001 µS/cm (minimum; varies with range) |
Conductivity accuracy | ±0.5% FS |
Reference temperature | 5, 10, 15, 18, 20, 25 °C (settable) |
EC calibration | Up to 5 points; reminder; auto recognition (10, 84, 500, 1413 µS/cm; 12.88 mS/cm; 146.5 µS/cm; 1.408 mS/cm; 12.85 mS/cm; 111.3 mS/cm) |
Resistivity | 5.00 Ω·cm – 100.0 MΩ·cm; 0.01 Ω·cm min; ±0.5% FS |
TDS | 0.000 ppm – 1000 ppt; 0.001 ppm min; ±0.5% FS |
Salinity | NaCl; 0.0 – 80.0 ppt; 0.1 ppt; ±1 ppt |
Temperature | −10 to 135 °C (14 to 275 °F); 0.1 °C resolution; ±0.1 °C accuracy; units °C/°F |
Reading modes | Auto Read (Fast/Medium/Slow), Timed, Continuous; prompts: Reading / Stable / Locked |
Compensation | ATC/MTC; types: none / linear / pure water; settable cell constant, temp. coefficient, TDS factor |
Data & GLP | 1000 groups; GLP features; log management |
I/O | USB 2.0 (flash, PC, scanner); RS-232 (printer, auto-sampler) |
Input connector | 5-pin aviation connector (EC + temperature probe) |
Display | Backlight: Yes; Auto shutdown: 1–60 min or off |
Protection | IP54 (splash-proof) |
Date & time | Yes |
Power | AC adapter 100–240 V input, DC 24 V output |
Dimensions | 220 × 195 × 68 mm |
Weight | 950 g (2.09 lb) |
Applications
Suitable for water quality testing, chemical analysis, food & beverage, pharmaceutical production, and academic research.
Multi-point calibration and reliable temperature compensation help ensure precise, repeatable results across ultra-low to high conductivity samples.
For SOP-compliant workflows, GLP logging and export via USB/RS-232 simplify traceability and audits while reducing rework.
Request a Quote
Need pricing, lead time, or accessories? Contact our sales team for a tailored offer for your lab.