Overview — EC510T handheld conductivity meter
EC510T handheld conductivity meter from COLO.Science is a rugged IP65 portable meter for EC, TDS, salinity, resistivity, and temperature. It adds a 4.3″ high-contrast color touchscreen, GLP-compliant memory (1000 groups), and USB 2.0 communication. Therefore, it suits water quality labs, environmental monitoring, and industrial process control.
General features — EC510T handheld conductivity meter
- 4.3″ color high-contrast LCD touchscreen.
- Intelligent OS: user, calibration, electrode, method, data and log management.
- Multi-reading: Auto-Read, Timed-Read, and Continuous-Read (Fast/Medium/Slow).
- Automatic/Manual temperature compensation (ATC/MTC) for accurate results.
- Auto-Hold senses and locks the endpoint.
- GLP data storage: 1000 sets for each parameter, with date & time and analysis tools.
- USB 2.0 communication (flash drive, PC, scanner).
- Auto power-off extends battery life; Reset restores factory defaults.
- IP65 waterproof for field and outdoor measurements.
- 1–5 point calibration with automatic standard recognition.
- Settable parameters: cell constant, reference temperature (5/10/15/18/20/25 °C), temp. coefficient, TDS factor.
- Temperature compensation types: none, linear, pure water.
What’s included
- DJS-1VTC EC electrode
- 1413 μS/cm EC standard solution, 50 ml, 1 vial
- Electrode holder
- Silicone rubber case
- Wristbands
- Touch pen
- Carrying case
Specifications — Model EC510T
Parameter Range Resolution Accuracy Notes Conductivity (EC) 0.000 μS/cm to 3000 mS/cm 0.001 μS/cm minimum (varies with range) ±0.5% FS Ref. temp 5/10/15/18/20/25 ºC; up to 5-point calibration; standards auto-recognized incl. 10, 84, 500, 1413 μS/cm; 12.88 mS/cm; 146.5 μS/cm; 1.408 mS/cm; 12.85 mS/cm; 111.3 mS/cm Resistivity 5.00 Ω·cm to 100.0 MΩ·cm 0.01 Ω·cm minimum ±0.5% FS Derived from EC per instrument settings TDS 0.000 ppm to 1000 ppt 0.001 mg/L minimum (varies with range) ±0.5% FS Configurable TDS factor Salinity 0.0 to 80.0 ppt 0.1 ppt ±1 ppt — Temperature −10 to 135 ºC, 14 to 275 ºF 0.1 ±0.1 ºC Units: ºC / ºF Measurement & data
- Reading modes: Auto-Read (Fast/Medium/Slow), Timed, Continuous
- Status prompts: Reading / Stable / Locked
- Temperature compensation: ATC and MTC; types: none / linear / pure water
- Calibration: up to 5 points with reminder; broad standard recognition
- Data storage: 1000 groups + on-meter data analysis (review/compare/recalculate)
Inputs, outputs & design — EC510T handheld conductivity meter
- Probe input: EC + temperature via 5-pin aviation connector
- Outputs: USB 2.0 (flash memory device, PC, scanner)
- Backlight: Yes
- Auto shutdown: 300, 600, 1200, 1800, 3600 s, or off
- IP rating: IP65
- Date & time: Yes
- Power: Rechargeable lithium battery; AC adapter (100–240 V input, DC 5 V output)
- Dimensions: 90 × 255 × 40 mm
- Weight: 500 g (1.1 lb)
- Carton (W×D×H): 410 × 330 × 165 mm; Gross weight: 5.5 kg
🔗 Helpful links
Measurement Parameters | Conductivity |
---|---|
Instrument Type | Benchtop |
Based on 0 reviews
Only logged in customers who have purchased this product may leave a review.